Model Number N/A
Device Problem Device Slipped (1584)
Patient Problem No Information (3190)
Event Date 04/17/2015
Event Type  Injury  
Event Description
It was reported that patient underwent a hip arthroscopy and labral repair on (b)(6) 2015.During the procedure, the suture anchors pulled out.Additional holes were drilled and two new suture anchors were utilized to complete the procedure.
 
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under warnings, number 5 states, "care is to be taken to ensure adequate soft tissue fixation at the time of surgery." this report is number 2 of 2 mdrs filed for the same event (reference 1825034-2015-02060 / 02061).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERKNOT LONG 1 #1 MB
Type of Device
FASTENER, FIXATION
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key4775668
MDR Text Key20810646
Report Number0001825034-2015-02061
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK110145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date01/31/2020
Device Model NumberN/A
Device Catalogue Number916030
Device Lot Number668750
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/21/2015
Initial Date FDA Received05/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-