The device referenced in this report was returned to olympus for evaluation.The evaluation confirmed a burning smell when the top cover was removed during inspection.The device was not able to be powered on and a visual inspection found a burnt out fuse.The burnt out fuse was replaced with a similar one and that fuse immediately burned out when the device was powered on.The device became inoperative due to a faulty converter (power unit).A test converter was used and the device functioned properly and resumed its normal operation and passed all the functional inspection.There was fluid invasion found underneath the front chassis, traces of moisture on the rear panel, and on the top cover.The cause of the reported complaint is most likely due to the fluid invasion which could have shorted the circuit power unit and blown the fuses.The instruction manual warns user: "to prevent operator shock or instrument damage keep fluids away from all electrical equipment.If fluids are spilled on or into the high flow insufflation unit, stop operation of the unit immediately.".
|