It was reported that during a laparoscopic liver procedure, when they went to torque the device, it would not tighten and became loose again.When the hand piece was inspected, there was a significant crack down the side.Another device was used to complete the procedure.There were no adverse consequences for the patient reported.
|
(b)(4).Nose cone the handpiece was received with the nose cone cracked and the mount was noted to be loose.No functional testing could be performed due to the nose cone was extremely cracked and no instrument could be attached to the handpiece.The instrument was disassembled to inspect internal components and the moisture indicator was positive.The transducer assembly was not held in place due to the cracked nose cone, so torquing on the disposable resulted in twisting only the handpiece transducer assembly until the internal wires got disconnected.Due to the cracked nose cone, moisture entered the hand piece mid housing a possible cause of the nose cone being cracked is the sterilization method due to the heating and cooling of the sterilization cycles is a stressor.It is possible that the ingress of moisture affected handpiece functionality.
|