• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ULTRACISION** HARMONIC SCALPEL** HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ULTRACISION** HARMONIC SCALPEL** HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054
Device Problems Crack (1135); Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/10/2014
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic liver procedure, when they went to torque the device, it would not tighten and became loose again.When the hand piece was inspected, there was a significant crack down the side.Another device was used to complete the procedure.There were no adverse consequences for the patient reported.
 
Manufacturer Narrative
(b)(4).Nose cone the handpiece was received with the nose cone cracked and the mount was noted to be loose.No functional testing could be performed due to the nose cone was extremely cracked and no instrument could be attached to the handpiece.The instrument was disassembled to inspect internal components and the moisture indicator was positive.The transducer assembly was not held in place due to the cracked nose cone, so torquing on the disposable resulted in twisting only the handpiece transducer assembly until the internal wires got disconnected.Due to the cracked nose cone, moisture entered the hand piece mid housing a possible cause of the nose cone being cracked is the sterilization method due to the heating and cooling of the sterilization cycles is a stressor.It is possible that the ingress of moisture affected handpiece functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4794858
MDR Text Key5878181
Report Number3005075853-2015-03195
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/21/2015
Initial Date FDA Received05/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR AND DISPOSABLE
-
-