• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION GENESIS PULSE GENERATOR, 8-CHANNEL; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION GENESIS PULSE GENERATOR, 8-CHANNEL; SCS IPG Back to Search Results
Model Number 3608
Device Problem Low Battery (2584)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 04/24/2015
Event Type  Injury  
Event Description
It was reported ((b)(6)) the patient observed an ipg low battery warning.It is unknown if the patient lost stimulation.The patient underwent surgical intervention where the ipg was replaced with a new one, which resolved the issue.Sjm was made aware of the issue on (b)(6) 2015, and the patient's scs system was not evaluated by an sjm representative to verify and/or troubleshoot any of the reported issues prior to explant.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Evaluation codes: results: the device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.Ipg battery pulse load testing confirmed the low battery message observed in the field was due to battery passivation.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENESIS PULSE GENERATOR, 8-CHANNEL
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4796561
MDR Text Key5835068
Report Number1627487-2015-10156
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2015
Device Model Number3608
Device Lot Number3978522
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/29/2015
Initial Date FDA Received05/26/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-