Model Number 3608 |
Device Problem
Low Battery (2584)
|
Patient Problem
Complaint, Ill-Defined (2331)
|
Event Date 04/24/2015 |
Event Type
Injury
|
Event Description
|
It was reported ((b)(6)) the patient observed an ipg low battery warning.It is unknown if the patient lost stimulation.The patient underwent surgical intervention where the ipg was replaced with a new one, which resolved the issue.Sjm was made aware of the issue on (b)(6) 2015, and the patient's scs system was not evaluated by an sjm representative to verify and/or troubleshoot any of the reported issues prior to explant.
|
|
Manufacturer Narrative
|
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Manufacturer Narrative
|
Evaluation codes: results: the device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.Ipg battery pulse load testing confirmed the low battery message observed in the field was due to battery passivation.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|