Evaluation summary: no data regarding product identity was received i.E.No lot or serial number were indicated for the event; therefore, the device history record (dhr) could not be reviewed.The sample returned on 14-may-2015, and sent for sterilization, therefore, the condition of the product was not verified yet.Because a sample was not investigated yet, the root cause cannot be determined.The root cause will be reassessed upon completing the investigation.Additional information was requested and received.(b)(4).
|
Additional information: evaluation summary: the device was returned for evaluation.During inner illumination the shunts' lumen was found to be partially blocked.After cleaning the shunt, the lumen was opened.There is no indication of manufacturing related factors that could cause the blockage.During production, 100% final inspection is being performed on the entire batch, including inner illumination.If such a defect had been noticed during the inspection, the product would have been rejected immediately.The shunt had been in the patient's eye.Having said that, one may conclude, that the blockage was formed after the product had left the manufacturing plant.The root cause cannot be conclusively determined since the blockage could have been caused by many different reasons during and after the clinical procedure.The blockage does not seem to be product related since the restriction unit was seen to be centered and after cleaning the device no light obstructions were seen.Therefore, there is no evidence for an inherent defect that might have caused the event, and the root cause cannot be conclusively determined.(b)(4).
|