Model Number 9207010-57065
Device Problems Crack (1135); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2015
Event Type  malfunction  
Event Description
The customer reported to customer service that the transformer housing of her advanced personal double breast pump was cracked open, exposing the underlying electronics, which is a safety risk.She also reported that the prongs had fallen out.
 
Manufacturer Narrative
The customer was sent a replacement power supply.The customer reported a breach in the transformer housing.She did not report of any fire, spark or injury.The product involved in the complaint was not returned for evaluation/investigation at this time.Therefore, no conclusions can be made as to the cause of the event.Should additional information of the original product be received, resulting in new, changed, or corrected information, a follow up report will be filed at that time.This issue with the damaged transformer housing for the pump in style device was addressed in investigation (b)(4).The investigation found that the transformers are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fail prematurely when subjected to normal use and forseeable misuse.The packaging used by the manufacturer to ship the transformers to medela is not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process has been modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength has also been increased to further protect the transformers during shipping.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BP ADVANCED PERSONAL DOUBLE
Manufacturer (Section D)
MEDELA, INC.
mchenry IL
Manufacturer Contact
don alexander
1101 corporate dr.
mchenry, IL 60050
8004358316
MDR Report Key4798311
MDR Text Key5831781
Report Number1419937-2015-00155
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/08/2015,05/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9207010-57065
Device Catalogue Number9207010-57065
Device Lot NumberREV L/UNK
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date05/08/2015
Initial Date Manufacturer Received 05/08/2015
Initial Date FDA Received05/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-