• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA INC. PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA INC. PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG Back to Search Results
Model Number 9207010/57036
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/18/2015
Event Type  malfunction  
Event Description
The customer reported to customer service that the transformer housing for her pump in style advanced breast pump was cracked and she can see inside it, which is a safety risk.
 
Manufacturer Narrative
The customer was sent a new transformer.The product involved in the complaint was not returned for eval/investigation at this time.Therefore, no conclusions can be made as to the cause of the event.Should additional info or the original product be received, resulting in new, changed, or corrected info, a follow up report will be filed at that time.This issue with the damaged transformer housing for the pump in style device was addressed in investigation (b)(4).The investigation found that the transformers are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fail prematurely when subjected to normal use and forseeable misuse.The packaging used by the manufacturer to ship the transformers to medela is not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process has been modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength has also been increased to further protect the transformers during shipping.Complaints against this product are currently being monitored for effectiveness of the above mentioned corrective action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG
Manufacturer (Section D)
MEDELA INC.
mchenry IL
Manufacturer Contact
don alexander
1101 corporate dr.
mchenry, IL 60050
8004358316
MDR Report Key4798313
MDR Text Key5879218
Report Number1419937-2015-00153
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9207010/57036
Device Catalogue Number9207010/57036
Device Lot NumberREV M-1212
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date05/18/2015
Initial Date Manufacturer Received 05/18/2015
Initial Date FDA Received05/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-