AMERICAN MEDICAL SYSTEMS (MN) AMS ACTICON NEOSPHINCTER; IMPLANTED FECAL INCONTINENCE DEVICE Back to Search Results
Device Problem Fluid/Blood Leak (1250)
Patient Problem Failure of Implant (1924)
Event Date 05/08/2015
Event Type  Injury  
Event Description
It was reported that the patient had his action neosphincter removed and replaced due to fluid loss, "cuff." there were no patient complications reported as a result of this event.
 
Manufacturer Narrative
(in order of cuff, pump, balloon): catalog #: 72401964, 72402287, 72402104; expiration date: 12/15/2015, 04/05/2014, 12/12/2017; serial #: (b)(4).(in order of cuff, pump, balloon): manufacture date: 6/26/2012, 04/09/2013, 01/02/2013.
 
Manufacturer Narrative
Device evaluation: the acticon device was visually inspected.There was a leak in the cuff at the face/backing junction that was due to fatigue.The pump and balloon were not tested due to the leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS ACTICON NEOSPHINCTER
Type of Device
IMPLANTED FECAL INCONTINENCE DEVICE
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
sharon zurn
10700 bren road w
minnetonka, MN 55343
9529306000
MDR Report Key4810950
MDR Text Key5922578
Report Number2183959-2015-00218
Device Sequence Number1
Product Code MIP
Combination Product (y/n)N
PMA/PMN Number
P010020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2015
Initial Date FDA Received06/02/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/02/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-