Catalog Number 09390116
Device Problem Material Fragmentation (1261)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/07/2015
Event Type  malfunction  
Event Description
The customer reported that during a thr procedure involving an exeter cemented stem, the exeter plug allegedly split on insertion.There was a delay to surgery of 5 minutes while the surgeon tried to retrieve the plug.Half of the plug was left in the patient and an additional plug was used to complete the procedure.
 
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Additional information was requested and if it becomes available will be submitted in a supplemental report.Product cannot be decontaminated.
 
Manufacturer Narrative
An event regarding a fractured device involving an exeter bone plug was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as the device was not returned.-medical records received and evaluation: not performed as no medical records were provided.-device history review: dhr review was satisfactory.-complaint history review: complaint history review confirmed that there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined with the available information.Return of the reported device for evaluation is required to complete the investigation for determining the root cause.A capa trend analysis was conducted for the reported failure mode and concluded that fracture of exeter bone plugs may result from factors such as errors made during use or design factors.No further investigation for this event is possible at this time as the device was not received by stryker orthopaedics.If the device becomes available, this investigation will be reopened.
 
Event Description
The customer reported that during a thr procedure involving an exeter cemented stem, the exeter plug allegedly split on insertion.There was a delay to surgery of 5 minutes while the surgeon tried to retrieve the plug.Half of the plug was left in the patient and an additional plug was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER 2.5 I M PLUG 16MM
Type of Device
IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
beverly lima
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4813065
MDR Text Key5917777
Report Number0002249697-2015-01780
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K980843
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/01/2019
Device Catalogue Number09390116
Device Lot NumberL7248
Other Device ID NumberSTERILE LOT# QA1S4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2015
Initial Date FDA Received06/02/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-