Catalog Number 09390116 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 05/07/2015 |
Event Type
malfunction
|
Event Description
|
The customer reported that during a thr procedure involving an exeter cemented stem, the exeter plug allegedly split on insertion.There was a delay to surgery of 5 minutes while the surgeon tried to retrieve the plug.Half of the plug was left in the patient and an additional plug was used to complete the procedure.
|
|
Manufacturer Narrative
|
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Additional information was requested and if it becomes available will be submitted in a supplemental report.Product cannot be decontaminated.
|
|
Manufacturer Narrative
|
An event regarding a fractured device involving an exeter bone plug was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as the device was not returned.-medical records received and evaluation: not performed as no medical records were provided.-device history review: dhr review was satisfactory.-complaint history review: complaint history review confirmed that there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined with the available information.Return of the reported device for evaluation is required to complete the investigation for determining the root cause.A capa trend analysis was conducted for the reported failure mode and concluded that fracture of exeter bone plugs may result from factors such as errors made during use or design factors.No further investigation for this event is possible at this time as the device was not received by stryker orthopaedics.If the device becomes available, this investigation will be reopened.
|
|
Event Description
|
The customer reported that during a thr procedure involving an exeter cemented stem, the exeter plug allegedly split on insertion.There was a delay to surgery of 5 minutes while the surgeon tried to retrieve the plug.Half of the plug was left in the patient and an additional plug was used to complete the procedure.
|
|