SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500 Back to Search Results
Catalog Number 10492730
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/13/2015
Event Type  malfunction  
Event Description
Customer reported that they noticed patient results showed up in the lis (data management system) did not match the results printed on the instrument.Customer indicated that although patient sample was run on (b)(6), the results that crossed were from a sample dated (b)(6) in the instrument.Customer stated that after this event, they have been manually editing the patient results crossing to the lis system.There was no report of injury due to this event.
 
Manufacturer Narrative
Siemens customer support engineer (cse) checked the instrument, backed up the unit setup and reinstalled the operating software.Cse verified the setup, ran quality control and two patients to verify proper communication with the lis.It passed.Reloading software resolved the customer's issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4827535
MDR Text Key5836670
Report Number1217157-2015-00072
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10492730
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/18/2015
Initial Date FDA Received06/09/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-