• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER PRIME? XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER PRIME? XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004EPTP4500THN40
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/06/2015
Event Type  malfunction  
Event Description
Reportable based on the product analysis completed on (b)(4) 2015.It was reported the during a flutter ablation procedure in the isthmus, he blazer prime¿ xp being used, kinked at the ablation tip.However, the returned product analysis revealed that the glue was lifting/peeling off electrodes on the curve.
 
Manufacturer Narrative
Patient age at time of event: 18 years or older.(b)(4).Device evaluated by manufacturer: a visual examination identified a kink at 13mm from the tip while in the neutral position.In addition the ring#2 has broken adhesive.X ray analysis reveled the center support is kink.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAZER PRIME? XP
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4828918
MDR Text Key19502273
Report Number2134265-2015-03681
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
PMA/PMN Number
P020025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/09/2016
Device Model NumberM004EPTP4500THN40
Device Catalogue NumberEPTP4500THN4
Device Lot Number16592566
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/14/2015
Initial Date FDA Received06/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-