• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS CORP PCS®2 PLASMA COLLECTION SYSTEM; SEPARATOR, AUTOMATED, BLOOD CELL, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORP PCS®2 PLASMA COLLECTION SYSTEM; SEPARATOR, AUTOMATED, BLOOD CELL, DIAGNOSTIC Back to Search Results
Model Number 06002-110-NA-EW
Device Problems Disconnection (1171); Difficult to Open or Close (2921)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 03/24/2014
Event Type  Injury  
Event Description
Haemonetics received a complaint on (b)(6) 2015 stating that a plasma center's employee reported that she strained her wrist when unlocking the centrifuge on the pcs®2 device.The employee's initial diagnosis in (b)(6) 2014 was for a wrist sprain, not otherwise specified.A later diagnosis was de quervain's tenosynovitis and epicondylitis of the elbow to a scaphoid lunate tear.This diagnosis led to surgical intervention with an extensor compartment release.The employee was on numerous work restrictions until being released from care on (b)(6) 2014.Haemonetics attempted to obtain additional information for this reported issue with no further information received.
 
Manufacturer Narrative
The customer could not provide a specific device serial number, however, provided a list of 18 possible devices that the employee was working on at the time.A haemonetics field service engineer completed a preventative maintenance check of each device on 06/09/2015.There were no issues found with the device cover locks.Haemonetics does not have evidence that there is a definitive causal relationship between the surgical intervention and the use of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PCS®2 PLASMA COLLECTION SYSTEM
Type of Device
SEPARATOR, AUTOMATED, BLOOD CELL, DIAGNOSTIC
Manufacturer (Section D)
HAEMONETICS CORP
400 wood rd
braintree MA 02184
Manufacturer (Section G)
HAEMONETICS CORP
400 wood rd
braintree MA 02184
Manufacturer Contact
julie smith, rn
400 wood rd
braintree, MA 02184
7819170643
MDR Report Key4837596
MDR Text Key5902245
Report Number1219343-2015-00007
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK060033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Service and Testing Personnel
Device Model Number06002-110-NA-EW
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/13/2015
Initial Date FDA Received06/11/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-