Model Number 100139 |
Device Problems
Incorrect Or Inadequate Test Results (2456); Low Test Results (2458); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Coagulation Disorder (1779); No Code Available (3191)
|
Event Date 05/19/2015 |
Event Type
Injury
|
Event Description
|
The customer called in questioning why they had not received the medical device correction notification that they had found on the alere website, while reviewing the return policy.The customer alleged variances between multiple inratio inr results and laboratory inr results on multiple patients with some patients admitted to the hospital.The specific dates and results were not available.There was only one (1) patient result provided; however, the date was unknown.Additionally, the specific lot number and monitor serial number could not be provided.The customer could not verify which of these lot numbers (357496 or 367327a) had the variance.Results are as follows: date: unknown, inratio inr: 3.1, laboratory inr: 12.6.Therapeutic range: unknown.The patient was hospitalized and outcome/treatment could not be provided by the customer.Though requested, there was no additional information provided.
|
|
Manufacturer Narrative
|
Estimated as the date the call was received, (b)(6) 2015, since the customer could not provide the actual date of the event and the call was received on (b)(6) 2015.The pt/inr test strips lot # 357496 referenced, is be reported under a separate manufacturer report number 2027969-2015-00406.Investigation pending.
|
|
Manufacturer Narrative
|
Investigation/conclusion: the products associated with the complaint were returned for investigation.A review of the monitor memories could not determine if either monitor was used to obtain the customer's reported result.The returned monitor memories were downloaded and reviewed.The reported strip lot #367327a has strip code# t87xk.The monitor's memory did not contain the reported strip code#t87xk.Should additional case information be provided in the future, further investigation with the appropriate monitor will be pursued.A review of in-house testing of lot 367327a was performed.The retain strip testing results met both accuracy and repeatability criteria.The product performed as expected and no product deficiencies were observed.The manufacturing records for the lot were reviewed and the lot met release specifications.The root cause could not be determined from the information provided by the customer.Based on the information available, there is no indication of a product deficiency and no corrective action is required.Further investigation was performed under capa-(b)(4).
|
|
Search Alerts/Recalls
|
|
|