CYBERONICS INC PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Device Displays Incorrect Message (2591); Difficult to Open or Close (2921)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2015
Event Type  malfunction  
Event Description
It was reported that the physician's tablet was giving an 'unable to open port' error message.The physician was provided a new usb cable and the issue resolved.The faulty usb cable was received for analysis.Analysis of the cable was completed on 06/11/2015.The cause for the reported allegation is associated with a disconnected wire connection in the returned serial cable.Once the wire was soldered onto the serial cable pcb, no further anomalies were identified.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4845286
MDR Text Key5946792
Report Number1644487-2015-04901
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 10.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 05/21/2015
Initial Date FDA Received06/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-