GE OEC MEDICAL SYSTEMS (SLC) 9800; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9800
Device Problems Device Stops Intermittently (1599); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2015
Event Type  malfunction  
Event Description
The customer reported that the system was shutting down without command of the operator.There is no report of a patient injury or death associated with this event.
 
Manufacturer Narrative
A ge service representative performed an onsite investigation.The 5 vdc power supply was evaluated and calibrated to the optimal operating voltage.The ac power cable connections were also repaired during the service event.The system was tested and found to be working as intended and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9800
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key4846899
MDR Text Key22152130
Report Number1720753-2015-02434
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9800
Device Lot Number82-0862-RC
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/21/2015
Initial Date FDA Received06/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-