• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/27/2015
Event Type  malfunction  
Event Description
The customer reported that the mean airway pressure was not stable while on a patient.The patient was removed from the oscillator and placed on another oscillator, no patient harm.
 
Manufacturer Narrative
(b)(4).The carefusion field service engineer evaluated the ventilator and found the mean airway pressure was reading lower than the flow analyzer and was fluctuating.Cleaned and reseated the connection to the pressure meter.Calibrated the airway pressure monitor transducer.Calibrated the pneumatics section after finding the bias flow was out of spec.Tightened loose nut and cross bolt on the min paw thumbwheel.Completed alarms checkout.Passed patient circuit calibration and vent performance check.No fluctuations and all readings within specifications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
75 n. fairway drive
vernon hills, IL 60061
8473628056
MDR Report Key4862846
MDR Text Key5808449
Report Number2021710-2015-01144
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/27/2015
Initial Date FDA Received06/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-