SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW STRATUS CS Back to Search Results
Catalog Number 10453531
Device Problem False Positive Result (1227)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/04/2015
Event Type  No Answer Provided  
Manufacturer Narrative
The cause for the discordant troponin result is unknown.Siemens is in process of investigating the issue.
 
Event Description
Customer reported false positive troponin results on the instrument.There was no report of injury due to this event.
 
Manufacturer Narrative
Siemens representative reviewed sample handling process with customer, quality control and calibration was passed.Message log was unavailable for review.Customer indicated that they have not had further issues.There is no evidence that a siemens product is not meeting specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATUS CS
Type of Device
STRATUS CS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
glasgow business community
newark DE 19702
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
glasgow business community
newark DE 19702
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4871173
MDR Text Key22997775
Report Number1217157-2015-00086
Device Sequence Number1
Product Code DAP
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K051650
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10453531
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/23/2015
Initial Date FDA Received06/26/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/23/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-