Catalog Number 155401135 |
Device Problems
Material Disintegration (1177); Insufficient Information (3190)
|
Patient Problems
Foreign Body Reaction (1868); No Code Available (3191)
|
Event Date 05/13/2010 |
Event Type
malfunction
|
Event Description
|
Litigation alleges patient suffers from metallosis and toxic cobalt chromium metal ions.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: unavailable.This complaint is the subject of litigation or a legal claim and currently complete product detail is not available at this time.A follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
Ppf alleges pseudotumor and metal wear.Added date of birth, surgeon, facility name and revision hospital.Updated unknown head, liner and stem.Corrected event date and date of revision.Doi: (b)(6) 2010; dor: (b)(6) 2010 (left hip).
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
Ppf alleges pseudotumor and metal wear.Added date of birth, surgeon, facility name and revision hospital.Updated unknown head, liner and stem.Corrected event date and date of revision.Doi: (b)(6) 2010; dor: (b)(6) 2010 (left hip).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|