(b)(4).Upon receipt at our post market quality assurance laboratory, visual inspection noted the helix was extended and dried blood was present in the helix housing.Resistance testing found the lead was not electrically continuous.Detailed analysis confirmed that the cathode conductor coil was fractured near the terminal pin.Microscopic analysis confirmed that the lead became fractured due to torsional overstress.A laboratory technician tested the helix mechanism and found it was nonfunctional, confirming the clinical observation.Based upon the clinical observations and the laboratory findings, we believe the conductor coil became fractured during attempts to extend/retract the helix.
|