MEDTRONIC HEART VALVES DIVISION ACCUTRAK DELIVERY CATHETER SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Model Number DCS-C4-18FR |
Device Problems
Positioning Failure (1158); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/09/2015 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that during the implant of a 26 mm transcatheter bioprosthetic valve, when the micro knob of this delivery catheter system (dcs) was turned to start the deployment of the valve, the capsule moved up to 2 mm and then it would not move any further.It was reported that there was no "clicking" heard.Per the instructions for use (ifu), upward pressure was applied, to assist with deployment.The capsule still would not move.The delivery catheter system (dcs) was taken out of the body and without doing anything differently, the valve was able to be deployed on the table.A new dcs was used to implant the valve.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Product analysis: the product has not been returned for analysis, however, the return is anticipated.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.Should the device be returned or additional information become available, a supplemental report will be submitted.(b)(4).
|
|
Manufacturer Narrative
|
Upon receipt at medtronic's quality laboratory, the delivery catheter system (dcs) was received tightly coiled in a plastic bag contained in an open biohazard bag.The dual shaft showed deformation which appeared to be the result of the receipt condition.A kink was observed on the dual shaft and appeared to be associated with the receipt condition.The handle appeared intact.The micro knob (thumb wheel) appeared to retract and advanced the capsule.The macro (cursor) moved to fully advance and retracted positions and locked in place when released.The capsule was able to fully advanced and retracted.The distal tip of the capsule seated flushed against the plunger tip when fully advanced.Two large kinks were observed on the distal and proximal ends of the capsule and were associated with the receipt condition.The distal end of the capsule appeared flattened with creases adjacent to and on the radiopaque marker band.This may be the result of retracting an unloaded capsule through the introducer.Minor buckling of the material was observed at the proximal end of the capsule.The plunger tube was bent at the hub transition with a small kink adjacent to the middle of the tube.Both are likely associated with the receipt condition.No gouge marks were observed on the catheter tabs.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|