SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; IMMUNOASSAY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR CP
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2015
Event Type  malfunction  
Event Description
Discordant results were obtained on patient samples tested for multiple methods on an advia centaur cp instrument.Discordant results were not reported to the physician(s), as the customer stated that results were not believable.Quality controls were out of range for all methods.It is unknown if repeat testing was performed.There are no reports of patient intervention or adverse health consequences due to the discordant results.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse replaced the wash 1 solution due to contamination and cleaned the luminometer due to base residue.The cause of the discordant results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR CP
Type of Device
IMMUNOASSAY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
STRATEC BIOMEDICAL SWITZERLAND AG
registration #: 3008494306
neuwiesenstrasse 4
beringen, 8222
SZ   8222
Manufacturer Contact
margarita karan
511 benedict avenue
tarrytown 10591
9145243105
MDR Report Key4894270
MDR Text Key6018866
Report Number2432235-2015-00318
Device Sequence Number1
Product Code MOI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR CP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/11/2015
Initial Date FDA Received07/07/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-