Model Number ADVIA CENTAUR CP |
Device Problem
Incorrect Or Inadequate Test Results (2456)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/11/2015 |
Event Type
malfunction
|
Event Description
|
Discordant results were obtained on patient samples tested for multiple methods on an advia centaur cp instrument.Discordant results were not reported to the physician(s), as the customer stated that results were not believable.Quality controls were out of range for all methods.It is unknown if repeat testing was performed.There are no reports of patient intervention or adverse health consequences due to the discordant results.
|
|
Manufacturer Narrative
|
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse replaced the wash 1 solution due to contamination and cleaned the luminometer due to base residue.The cause of the discordant results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
|