Complaint received reporting reading issues with use of one 52510-15, 8f optiq, svo2/cco catheter.The initial information received reported "thermal coil fault, switched cables and q2+.Catheter would not read cco/cci.Svo2 worked fine." it was reported that this issue did contribute a delay as attending clinicians were troubleshooting the devices, equipment and changeouts.The 52510-15 catheter device was removed and replaced with no further issues encountered.The catheter was successfully pre-tested during initial set-ups as well as pre-insertion tested, successfully placed and in use for approximately five hours into the procedure when the reading issues occurred.There were no reported patient injuries and or adverse consequences.Device return: one used 52510-15 catheter, unknown accessory tubing devices attached to the dtp, ra and pa lines.
|