• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ULTRACISION HARMONIC ACE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ULTRACISION HARMONIC ACE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number ACE36E
Device Problems Break (1069); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2015
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure, the device could not be activated.The titanium tip broke after re-installation.Another like device was used to complete the procedure.There were no adverse consequences for the patient reported.
 
Manufacturer Narrative
(b)(4) no device received for analysis at time of submission of 3500a.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.Additional information received: did the broken piece fall into the patient? no.Is the broken piece returning with the device or was it disposed of? the broken blade tip was discarded.
 
Manufacturer Narrative
(b)(4).Additional information: batch # l9283v.The device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched, evidence of contact with metal in or out of the operative field.The device was functionally tested with a generator.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.The batch history records were reviewed with no anomalies noted during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRACISION HARMONIC ACE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
milton garrett
5133378865
MDR Report Key4894930
MDR Text Key16990725
Report Number3005075853-2015-04083
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K042777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/15/2019
Device Catalogue NumberACE36E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/22/2015
Initial Date FDA Received07/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/03/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR AND HANDPIECE
-
-