Model Number 01-SETSCREW |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problems
Intraoperative Pain (2662); No Information (3190)
|
Event Date 03/24/2015 |
Event Type
Injury
|
Event Description
|
Patient had a two level lumbar fusion surgery on (b)(4) 2015.On (b)(4) 2015 it was identified in a follow up visit that one of the locking set screws had come off the construct.The patient was revised on (b)(4) 2015.
|
|
Manufacturer Narrative
|
During the final tightening of the set screws the surgeon used the compressor.The surgical technique for this system specifically states "do not attempt to apply full locking torque to the set screw with the compressor in place.Remove the compressor and final driver, then proceed to final set screw locking." this may have prevented the set screw from receiving the correct amount of torque for final locking.These devices have not been returned for inspection.
|
|
Manufacturer Narrative
|
The devices have been returned for inspection.All devices were found to be within specification.After further review there were 4 set screws in the construct that appear to have not been final locked down.The correct technique for final locking of the set screws is defined in this system's surgical technique.
|
|
Event Description
|
Patient was revised on (b)(6) 2015.There were 4 loose setscrews found during this revision.The rods, set screws and pedicle screws were explanted and replaced successfully.
|
|
Search Alerts/Recalls
|