• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T165
Device Problems Over-Sensing (1438); Pacing Problem (1439); Device Sensing Problem (2917)
Patient Problem Cardiac Arrest (1762)
Event Date 06/23/2015
Event Type  malfunction  
Event Description
Boston scientific received information that this implantable cardioverter defibrillator (icd) oversensed noise on the right atrial (ra) lead and right ventricular (rv) lead channels resulting in pacing inhibition greater than 2 seconds in duration.Boston scientific technical services (ts) noted that the noise appears myopotential in nature and suspects lead insulation damage possibly from lead-on-lead contact.The device was reprogrammed and remains in service.No adverse patient effects were reported.
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4913720
MDR Text Key6051522
Report Number2124215-2015-09117
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/05/2010
Device Model NumberT165
Other Device ID NumberVITALITY 2 DR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/23/2015
Initial Date FDA Received07/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4096; 0185; T165
-
-