• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUNCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUNCY Back to Search Results
Model Number 3100A
Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2015
Event Type  malfunction  
Event Description
The customer reported that while on a patient, the mean airway pressure started "drifting all over the place" so much that they didn't want to use it any more.No patient compromise.
 
Manufacturer Narrative
(b)(4).The third party service company evaluated the ventilator and was unable to verify the reported complaint.They found that the mean airway pressure was stable and the performance inspection passed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUNCY
Manufacturer (Section D)
CAREFUSION
yorba linda CA
Manufacturer (Section G)
CAREFUSION
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
75 n. fairway drive
vernon hills, IL 60061
8473628056
MDR Report Key4918232
MDR Text Key6026667
Report Number2021710-2015-01206
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901-RNT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2015
Initial Date FDA Received07/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-