• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) ATLANTIS? SR PRO; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) ATLANTIS? SR PRO; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number H749389420
Device Problems Retraction Problem (1536); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
Same case as: 2134265-2015-04595 and 2134265-2015-04543.It was reported that automatic pullback failure occurred.During percutaneous coronary intervention (pci), an ilab ultrasound imaging system was used in conjunction with an atlantis¿ sr pro imaging catheter and a sled.It was noted that after connecting the motor drive unit (mdu) to the galaxy sled and pressing the pullback record button, an error message appeared and the sled would not pullback.The procedure was completed through manual pullback.The patient's status is stable and fine.
 
Manufacturer Narrative
Age at time of event: 18 years or older.Device evaluated by mfr: the complaint device was not returned by the customer; therefore, no product analysis could be performed.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause was unable to be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLANTIS? SR PRO
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4920012
MDR Text Key6010626
Report Number2134265-2015-04612
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K063312
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749389420
Device Catalogue Number38942
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2015
Initial Date FDA Received07/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-