Model Number PT100
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
A customer in (b)(6) reported that the audible alarm on their pt100 myairvo humidifier was not working.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint myairvo humidifier was returned to fisher & paykel healthcare in (b)(4) and was visually inspected and electrically tested.Results: during testing the airvo functioned normally.The visual alerts were operating, however no audible alarm was heard.The fault was traced to a faulty speaker and electrical resistance testing showed that the speaker's resistance was open circuit.A lot check revealed no other complaints of this nature for lot number 141119.Conclusion: the supplier of the speaker unit was notified and they have carried out an investigation.The problem has been traced to an issue with the gluing process.The supplier has adjusted the process and taken steps to ensure that each speaker is checked following the gluing process and that any found faulty are discarded.The myairvo user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases." and that "the unit is not intended for life support." the user manual also contains instructions on how to check the alarm system functionality and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYAIRVO HUMIDIFIER
Type of Device
MYAIRVO HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9494534000
MDR Report Key4920392
MDR Text Key6027192
Report Number9611451-2015-00315
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPT100
Device Catalogue NumberPT100
Device Lot Number141119
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/18/2015
Initial Date FDA Received07/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-