• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4711K
Device Problem Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 07/01/2015
Event Type  malfunction  
Event Description
It was reported that small volume folfusor had a melted seam which connected the ¿flow rate connector to the line¿.There was no patient involvement as this was noticed before patient use.No additional information is available.
 
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).This lot was manufactured from march 18, 2015 to march 19, 2015 the device was received for evaluation.During visual inspection, the two sides of the flow restrictor housing were observed to be pinched inward along with an indentation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4924984
MDR Text Key6049182
Report Number1416980-2015-29407
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Other
Type of Report Initial
Report Date 07/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/01/2017
Device Catalogue Number2C4711K
Device Lot Number15C038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/02/2015
Initial Date FDA Received07/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-