• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750180
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/24/2015
Event Type  malfunction  
Manufacturer Narrative
A service visit was performed.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Additional information has been requested but not received to date.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during surgery the instrument froze/blocked.A system message was displayed.After re-boot it worked properly and surgery was successfully completed without any delay.There was no patient harm reported.Additional information has been requested but not received to date.
 
Manufacturer Narrative
Evaluation summary: the system was examined by a company representative and the reported event was not replicated.Several parts (footswitch, power distribution printed circuit board (pcb), supply cable, communication distribution pcb, footswitch interface pcb, footswitch cable, cabling and the direct current (dc) power cable) were all replaced to resolve the issue.The system was tested and found to meet product specifications.The system met specifications at the time of release.The company representative found an issue with the system, but replaced numerous parts at the same time.Therefore, the root cause of the reported event cannot be determined conclusively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key4927390
MDR Text Key22561479
Report Number2028159-2015-07132
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Remedial Action Other
Report Date 09/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750180
Other Device ID Number2.06
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/21/2015
Date Device Manufactured04/08/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-