Evaluation summary: the system was examined by a company representative and the reported event was not replicated.Several parts (footswitch, power distribution printed circuit board (pcb), supply cable, communication distribution pcb, footswitch interface pcb, footswitch cable, cabling and the direct current (dc) power cable) were all replaced to resolve the issue.The system was tested and found to meet product specifications.The system met specifications at the time of release.The company representative found an issue with the system, but replaced numerous parts at the same time.Therefore, the root cause of the reported event cannot be determined conclusively.
|