SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; ILLUMINATOR, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Catalog Number 72200568
Device Problem Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/03/2013
Event Type  malfunction  
Manufacturer Narrative
The device was returned for evaluation where product investigation revealed no burned parts found inside the unit.The product passed functional testing during two 8 hour burn-in periods inside the test tower.No burning smell was noted and all functions performed as expected.No problem was found.No further investigation is necessary.(b)(4).
 
Event Description
It was reported that when turning the device on before starting the case, the unit smoked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE, 500XL, XENON
Type of Device
ILLUMINATOR, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
150 minuteman road
andover, MA 01810
5123585706
MDR Report Key4935361
MDR Text Key22777702
Report Number1643264-2015-00056
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Report Date 07/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2013
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-