Model Number 801763
Device Problems Failure to Run on Battery (1466); Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 07/01/2015
Event Type  malfunction  
Event Description
It was reported that during the use of the device for a non-clinical activity, when the system-1 was unplugged, the system had no battery back-up.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.The field service representative (fsr) walked the biomedical engineer (biomed) through proper battery charging and the biomed confirmed it was at full charge the next day.Before, they had only plugged the system 1 into the wall socket, but not turned it on.The fsr explained to them proper maintenance.No additional action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
APS 1 (HEART LUNG CONSOLE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4938385
MDR Text Key6193376
Report Number1828100-2015-00611
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,health professional,use
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/01/2015
Initial Date FDA Received07/22/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-03/13/12-004-C
Patient Sequence Number1
-
-