• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK ROTALINK? PLUS; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK ROTALINK? PLUS; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H749236310020
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported that foreign material was found on the burr.A 1.25mm rotalink¿ plus was selected for use.During preparation, while setting the speed, the burr was advanced forward; however, it was noted that the burr tangled up in a towel.Pieces of the towel and lint were observed on the burr.The burr was not placed in the patient.The procedure was completed with another of the same device.No patient complications were reported.
 
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause has been determined to be use/user error as the dfu states: ¿the burr at the distal tip of the rotalink catheter is capable of rotating at very high speed.Do not allow parts of the body or clothing to come in contact with the burr.Contact may result in physical injury or entanglement.¿ (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK? PLUS
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4940726
MDR Text Key6253662
Report Number2134265-2015-04778
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749236310020
Device Catalogue Number23631-002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2015
Initial Date FDA Received07/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-