• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 70104.3262
Device Problems Out-Of-Box Failure (2311); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 08/15/2013
Event Type  malfunction  
Event Description
On (b)(6) 2013 at the (b)(6) center it was reported that the dpm 20-440 dual pressure module serial number unknown used with the hl 20 5 pumps console base serial number (b)(4) did not detect channel 2 pressure.This is an out of box failure/no patient involvement.(b)(4).
 
Manufacturer Narrative
Maquet medical system, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).The dpm was investigated at the ssu and a loose pin was found in the channel 2 transducer socket.The loose pin was placed back into the socket and secured using a nylon locking cable tie.The issue was resolved.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
kehler strasse 31
rastatt, 7643 7
GM  76437
Manufacturer (Section G)
MAQUET CARDIOPULMONARY AG
kehler strasse 31
rastatt, 7643 7
GM   76437
Manufacturer Contact
tina evancho
45 barbour pond drive
wayne, NJ 07470
9737097265
MDR Report Key4944655
MDR Text Key6256263
Report Number8010762-2014-00897
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/15/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number70104.3262
Device Catalogue NumberMCP00706195
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/15/2013
Device Age1 MO
Event Location Hospital
Date Report to Manufacturer08/15/2013
Initial Date Manufacturer Received 08/15/2013
Initial Date FDA Received07/28/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-