• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+ Back to Search Results
Catalog Number 10379675
Device Problems Nonstandard Device (1420); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/30/2015
Event Type  malfunction  
Manufacturer Narrative
Customer has been instructed to discard the power supply according to local regulations and not to use it again.An ufsn was provided to the customer.Customer confirmed that they were not adversely affected by defective power supply.Customer will be provided with a new power supply.The original equipment manufacturer (oem) of these power supplies has identified following root causes for the damaged power supply adaptors.1.The incorrect torque setting was selected by the production operator for the screws used to secure the power supply adaptor housings together.2.Mineral oil was found on the screws used to secure the housings together.The oil is used during the machining process during the production of the screws and was not removed by the screw supplier.These two root causes resulted in a combination of mechanical and chemical stress which led to the cracking of the screw pillars inside the housing top of the power supply adaptor.Manufacturing and inspection procedures have been revised at the manufacturer of the power supplies to address this issue.These corrective actions were finalized and implemented in june 2014.Siemens issued an urgent field safety notice ((b)(4)) in september 2014 to notify all affected siemens customers.
 
Event Description
Customer reported that they received defective power supply.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CT STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4950416
MDR Text Key24651738
Report Number1217157-2015-00109
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Assistant
Remedial Action Recall
Report Date 07/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/29/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0081-2015
Patient Sequence Number1
-
-