Model Number MODEL 250
Device Problems Device Displays Incorrect Message (2591); Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/30/2015
Event Type  malfunction  
Manufacturer Narrative
Suspect device udi: (b)(4).
 
Event Description
It was reported that the tablet was giving the error message "unable to open port".The site had a backup programming system, so there were no patients affected.Troubleshooting was attempted, but was unable to resolve the issue.Another serial data cable was used with the tablet and there were no further issues.The serial data cable was received, but analysis has not been completed to date.
 
Event Description
Product analysis was completed and it was reported that usb serial cable for the tablet device was broken.The cause for the reported allegation is associated with a disconnected wire connection in the returned serial cable.Once the wire was soldered onto the serial cable pcb, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4950867
MDR Text Key23130405
Report Number1644487-2015-05364
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Report Date 06/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 10.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2015
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-