• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. MULTIFIRE FASTPASS SCORP; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. MULTIFIRE FASTPASS SCORP; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE Back to Search Results
Catalog Number AR-13997MF
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 07/15/2015
Event Type  Injury  
Event Description
It was reported that the event occurred during the repair procedure.While pulling the tendon laterally, the tip of the multifire fastpass scorpion broke in the soft part of the patient.They tried to retrieve the broken piece but it is stuck and lost in the soft tissue.Surgery was finished without any further incident.
 
Manufacturer Narrative
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested/is expected for evaluation but has not yet been received.The cause of the event could not be determined from the information available and without device evaluation.Lot number was not provided so device history record review cannot be performed.This type of event is typically caused by user mechanical damage from dropping the device or hitting the device against bone or another device.The potential cause(s) of this event will be communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.Device expected but not yet returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE FASTPASS SCORP
Type of Device
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key4953905
MDR Text Key20808972
Report Number1220246-2015-00201
Device Sequence Number1
Product Code MDM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberAR-13997MF
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/16/2015
Initial Date FDA Received07/30/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-