|
Model Number HERO 1001 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Abscess (1690); Bacterial Infection (1735); Hematoma (1884); Pseudoaneurysm (2605)
|
Event Date 03/06/2014 |
Event Type
Injury
|
Manufacturer Narrative
|
Note: this mdr is being re-sent to mark type of report as initial.The ifu also lists pseudoaneurysm and seromas as potential vascular graft and catheter complications.Seven pseudoaneurysm cases were reported in 3 patients.At the time of pseudoaneurysm identification, grafts were cannulated.As stated in the ifu, rotation of cannulation sites is needed to avoid pseudoaneurysm formation.Details of dialysis sessions were unknown and compliance with rotation cannot be confirmed.The root cause for the reported event is unknown; however, all complications noted in the complaint are known potential complication of the hero graft.The ifu lists the following potential complications with the use of the hero graft: seroma, infection, vascular graft revision/replacement, partial stenosis or full occlusion of prosthesis or vasculature, pseudoaneurysm, hematoma, and abnormal healing.The hero graft is unlikely to be the direct source of the infection as the product undergoes a validated terminal sterilization process.There is no indication that an error or deficiency occurred at cryolife and the ifu adequately communicates risk.
|
|
Event Description
|
According to the publication, experience of hero dialysis graft placement in a challenging population, a total of 11 patients underwent 12 hero graft implants.All of the patients had cvod to varying extents, including subclavian vein stenosis or occlusion, innominate vein stenosis and superior vena cava stenosis.All of these patients had multiple vascular accesses (catheters and av fistulas or grafts) placed prior to the hero, with a minimum of 2 and a maximum of >14.Only 3 patients had a history of bacteremia, leading to removal of their access prior to hero placement.Reasons for reintervention or failure (n=11): thrombosis (total of 5 events, number of failed grafts 4), local infection (total of 3 events, number of failed grafts 3), pseudoaneurysm (total of 2 events), seroma (total of 1 event), hematoma (total of 1 event) and nonhealing incision (total of 1 event).Associated with the local infection the paper added the following information "grafts removed due to abscess formation over graft." additional information was received from the surgeon which showed that patient 8 had experienced a pseudoaneurysm after the original publication was printed.Hero 1001 and hero 1002 were investigated.Patient 8 was a male who had a hero graft (lot numbers unknown) implanted on (b)(6) 2010 and was explanted on (b)(6) 2014 due to pseudoaneurysm, hematomas and infection.He had a second hero graft (lot numbers unknown) implanted on (b)(6) 2014.The patient had a mechanical thrombectomy and revision of graft for thrombosis on (b)(6) 2010.A local infection was identified on (b)(6) 2014.The hero graft was not being cannulated during infection and there are no culture results.Three pseudoaneurysms were noted, two in the original hero graft and one in the subsequent hero graft.The pseudoaneurysms were identified on (b)(6) 2010, (b)(6) 2013 and (b)(6) 2014; the hero graft was being cannulated during the pseudoaneurysm in all three cases.Bacteremia was confirmed in the patient on (b)(6) 2014 and the graft was not cannulated during the bacteremia.Blood culture result showed gram negative bacilli.Two seromas were identified on (b)(6) 2014 and (b)(6) 2015; however, location of seromas is unknown.A total of five hematomas are listed occurring in both hero grafts.A hematoma was identified on (b)(6) 2010 occurring in the right arm proximal to mid upper arm and the hero was being cannulated during the hematoma.The second hematoma was identified on (b)(6) 2013 occurring in the right mid upper arm and the hero graft was cannulated during the hematoma.The third hematoma was identified on (b)(6) 2014 occurring in the right arm distal upper arm and right arm extending from proximal upper arm to the shoulder and the graft was not cannulated during the hematoma.The fourth and fifth hematomas were identified on (b)(6) 2014 occurring in right arm proximal to mid upper arm.Each event which occurred in patient 8 was investigated and a separate medwatch filed.This medwatch is filed for the pseudoaneurysm which occurred on (b)(6) 2014.
|
|
Manufacturer Narrative
|
According to the publication, experience of hero dialysis graft placement in a challenging population, a total of 11 patients underwent 12 hero graft implants.All of the patients had cvod to varying extents, including subclavian vein stenosis or occlusion, innominate vein stenosis and superior vena cava stenosis.All of these patients had multiple vascular accesses (catheters and av fistulas or grafts) placed prior to the hero, with a minimum of 2 and a maximum of >14.Only 3 patients had a history of bacteremia, leading to removal of their access prior to hero placement.Reasons for reintervention or failure (n=11): thrombosis (total of 5 events, number of failed grafts 4), local infection (total of 3 events, number of failed grafts 3), pseudoaneurysm (total of 2 events), seroma (total of 1 event), hematoma (total of 1 event) and nonhealing incision (total of 1 event).Associated with the local infection the paper added the following information "grafts removed due to abcess formation over graft." additional information was received from the surgeon which showed that patient 8 had experienced a pseudoaneurysm after the original publication was printed.Hero 1001 and hero 1002 were investigated.Patient 8 was a male who had a hero graft (lot numbers unknown) implanted on (b)(6) 2010 and was explanted on (b)(6) 2014 due to pseudoaneurysm, hematomas and infection.He had a second hero graft (lot numbers unknown) implanted on (b)(6) 2014.The patient had a mechanical thrombectomy and revision of graft for thrombosis on (b)(6) 2010.A local infection was identified on (b)(6) 2014.The hero graft was not being cannulated during infection and there are no culture results.Three pseudoaneurysms were noted, two in the original hero graft and one in the subsequent hero graft.The pseudoaneurysms were identified on (b)(6) 2010, (b)(6) 2013 and (b)(6) 2014; the hero graft was being cannulated during the pseudoaneurysm in all three cases.Bacteremia was confirmed in the patient on (b)(6) 2014 and the graft was not cannulated during the bacteremia.Blood culture result showed gram negative bacilli.Two seromas were identified on (b)(6) 2014 and (b)(6) 2015; however, location of seromas is unknown.A total of five hematomas are listed occurring in both hero grafts.A hematoma was identified on (b)(6) 2010 occurring in the right arm proximal to mid upper arm and the hero was being cannulated during the hematoma.The second hematoma was identified on (b)(6) 2013 occurring in the right mid upper arm and the hero graft was cannulated during the hematoma.The third hematoma was identified on (b)(6) 2014 occurring in the right arm distal upper arm and right arm extending from proximal upper arm to the shoulder and the graft was not cannulated during the hematoma.The fourth and fifth hematomas were identified on (b)(6) 2014 occurring in right arm proximal to mid upper arm.Each event which occurred in patient 8 was investigated and a separate medwatch filed.This medwatch is filed for the pseudoaneurysm which occurred on (b)(6) 2014.The ifu also lists pseudoaneurysm and seromas as potential vascular graft and catheter complications.Seven pseudoaneurysm cases were reported in 3 patients.At the time of pseudoaneurysm identification, grafts were cannulated.As stated in the ifu, rotation of cannulation sites is needed to avoid pseudoaneurysm formation.Details of dialysis sessions were unknown and compliance with rotation cannot be confirmed.The root cause for the reported event is unknown; however, all complications noted in the complaint are known potential complication of the hero graft.The ifu lists the following potential complications with the use of the hero graft: seroma, infection, vascular graft revision/replacement, partial stenosis or full occlusion of prosthesis or vasculature, pseudoaneurysm, hematoma, and abnormal healing.The hero graft is unlikely to be the direct source of the infection as the product undergoes a validated terminal sterilization process.There is no indication that an error or deficiency occurred at cryolife and the ifu adequately communicates risk.[(b)(4)].
|
|
Event Description
|
According to the publication, experience of hero dialysis graft placement in a challenging population, a total of 11 patients underwent 12 hero graft implants.All of the patients had cvod to varying extents, including subclavian vein stenosis or occlusion, innominate vein stenosis and superior vena cava stenosis.All of these patients had multiple vascular accesses (catheters and av fistulas or grafts) placed prior to the hero, with a minimum of 2 and a maximum of >14.Only 3 patients had a history of bacteremia, leading to removal of their access prior to hero placement.Reasons for reintervention or failure (n=11): thrombosis (total of 5 events, number of failed grafts 4), local infection (total of 3 events, number of failed grafts 3), pseudoaneurysm (total of 2 events), seroma (total of 1 event), hematoma (total of 1 event) and nonhealing incision (total of 1 event).Associated with the local infection the paper added the following information "grafts removed due to abcess formation over graft." additional information was received from the surgeon which showed that patient 8 had experienced a pseudoaneurysm after the original publication was printed.Hero 1001 and hero 1002 were investigated.Patient 8 was a male who had a hero graft (lot numbers unknown) implanted on (b)(6) 2010 and was explanted on (b)(6) 2014 due to pseudoaneurysm, hematomas and infection.He had a second hero graft (lot numbers unknown) implanted on (b)(6) 2014.The patient had a mechanical thrombectomy and revision of graft for thrombosis on (b)(6) 2010.A local infection was identified on (b)(6) 2014.The hero graft was not being cannulated during infection and there are no culture results.Three pseudoaneurysms were noted, two in the original hero graft and one in the subsequent hero graft.The pseudoaneurysms were identified on (b)(6) 2010, (b)(6) 2013 and (b)(6) 2014; the hero graft was being cannulated during the pseudoaneurysm in all three cases.Bacteremia was confirmed in the patient on (b)(6) 2014 and the graft was not cannulated during the bacteremia.Blood culture result showed gram negative bacilli.Two seromas were identified on (b)(6) 2014 and (b)(6) 2015; however, location of seromas is unknown.A total of five hematomas are listed occurring in both hero grafts.A hematoma was identified on (b)(6) 2010 occurring in the right arm proximal to mid upper arm and the hero was being cannulated during the hematoma.The second hematoma was identified on (b)(6) 2013 occurring in the right mid upper arm and the hero graft was cannulated during the hematoma.The third hematoma was identified on (b)(6) 2014 occurring in the right arm distal upper arm and right arm extending from proximal upper arm to the shoulder and the graft was not cannulated during the hematoma.The fourth and fifth hematomas were identified on (b)(6) 2014 occurring in right arm proximal to mid upper arm.Each event which occurred in patient 8 was investigated and a separate medwatch filed.This medwatch is filed for the pseudoaneurysm which occurred on (b)(6) 2014.
|
|
Search Alerts/Recalls
|
|
|