Model Number 60-00-60 |
Device Problems
Loose or Intermittent Connection (1371); Pumping Stopped (1503)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/03/2015 |
Event Type
malfunction
|
Event Description
|
Sorin group (b)(4) received a report the centrifugal pump stopped during a procedure.Dashes were shown on the display.There was no report of patient injury.Intermittent failure.
|
|
Manufacturer Narrative
|
Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the centrifugal pump stopped during a procedure.Dashes were shown on the display.There was no report of patient injury.The investigation is ongoing.A follow-up will be sent when the investigation is complete.
|
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was unable to reproduce the reported fault.The cp5 was removed from service and returned to livanova (b)(4) for further investigation.Visual inspection of the device at livanova (b)(4) did not identify any abnormalities or defects.Functional testing was able to reproduce the reported issue.The flow board was identified as faulty and was replaced to resolve the issue.The device was tested for 24 hours without further failures.The device was returned to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
|