SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5) ; CARDIOPULMONARY BYPASS PUMP SPEED CONTROL Back to Search Results
Model Number 60-00-60
Device Problems Loose or Intermittent Connection (1371); Pumping Stopped (1503)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/03/2015
Event Type  malfunction  
Event Description
Sorin group (b)(4) received a report the centrifugal pump stopped during a procedure.Dashes were shown on the display.There was no report of patient injury.Intermittent failure.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the centrifugal pump stopped during a procedure.Dashes were shown on the display.There was no report of patient injury.The investigation is ongoing.A follow-up will be sent when the investigation is complete.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was unable to reproduce the reported fault.The cp5 was removed from service and returned to livanova (b)(4) for further investigation.Visual inspection of the device at livanova (b)(4) did not identify any abnormalities or defects.Functional testing was able to reproduce the reported issue.The flow board was identified as faulty and was replaced to resolve the issue.The device was tested for 24 hours without further failures.The device was returned to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CARDIOPULMONARY BYPASS PUMP SPEED CONTROL
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key4960905
MDR Text Key6252532
Report Number9611109-2015-00266
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-00-60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/07/2015
Initial Date FDA Received07/29/2015
Supplement Dates Manufacturer Received02/19/2016
Supplement Dates FDA Received08/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-