• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/14/2015
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6), 2015.According to the complainant, during unpacking, the bristled portion of the brush was already bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Manufacturer Narrative
The complainant was unable to report the lot number; therefore the manufacture date and expiration date are unknown.However, the complainant stated that the device was used prior to the expiration date.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Manufacturer Narrative
Visual analysis of the returned rx cytology brush found the working length including catheter and pull wire, was bent in multiple places and kinked at 62.5 cm from the heat shrink.Assessment of pull wire at the brush bristle section found the wire had been bent at 1.3 cm from distal tip.The pull wire distal end lumen had been torn for about 2 mm.The brush was bent, functional evaluation found it would fully extend without issue and could not retract into the catheter because of the bent wire.It is possible that the complaint device pull wire was bent due to some aspect of device handling, possibly during shipping, storage, unpacking, or preparation.Therefore, the most probable root cause for this event is determined to be handling damage.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2015.According to the complainant, during unpacking, the bristled portion of the brush was already bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4975678
MDR Text Key18723957
Report Number3005099803-2015-02216
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,company represent
Reporter Occupation Other
Type of Report Initial
Report Date 07/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberM00545000
Device Catalogue Number4500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/15/2015
Initial Date FDA Received08/05/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-