No discrepant high results were observed from any devices tested with normal whole blood (wb) donors.Data results from all devices tested produced the same clinical outcome.Reviewed the manufacturing batch records for d-dimer lot w60220b.No issues with d-dimer recovery observed.No non-conformances were generated on the lot during production.Lot passed all final release specifications.As of (b)(6) 2015, this is the only complaint against d-dimer lot w60220b.No sample was returned.Unable to rule out sample specific interference as a potential cause for discrepant results.No product deficiency was established with the triage d-dimer device lot.No corrective action required at this time.
|