Model Number 8700-0752-01 |
Device Problems
Failure to Charge (1085); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/22/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product in complaint was returned to zoll on (b)(6) 2015 for investigation.However, investigation is still in progress.A supplemental report will be filed once investigation has been completed.
|
|
Event Description
|
It was reported that the autopulse li-ion battery (b)(4) failed.There was no report of any patient involvement.Please note that manufacturer has requested additional information, however no further details have been obtained.
|
|
Manufacturer Narrative
|
The li-ion battery (s/n (b)(4)) was returned to zoll (b)(4) for evaluation.Visual inspection was performed and no physical damage was observed to the battery.Functional testing of the battery was performed and the battery would not charge.As a result, the data archive could not be retrieved for review.The reported complaint of the battery has failed was confirmed.Since the battery archive could not be retrieved, therefore a root cause could not be determined.
|
|