Additional suspect medical device components involved in the event: model #: sc-2316-70, serial/lot #: (b)(4), description: infinion 70 cm lead kit; model #: sc-3400-30, serial/lot #: (b)(4), description: infinion splitter 2x8 kit (30 cm); model #: sc-4316, serial/lot #: (b)(4), description: next generation anchor kit-sterile.Sc-2316-70 (sn (b)(4)) the complaint was confirmed.The infinion lead body was fractured at the bent/kinked section of the lead body, 1 cm from the click anchor set screw marks.There were no exposed cables.There are 10 cable fractures.Sc-2316-70 (sn (b)(4)) the complaint was confirmed.The infinion lead body was fractured at the bent/kinked section of the lead body, 1 cm from the click anchor set screw marks.There were no exposed cables.There are 12 cable fractures.Sc- 3400-30 (sn (b)(4)) device evaluation indicated that the splitter passed visual, electrical and photographic imaging tests performed.Both splitters exhibited no anomalies.Sc-4316 (ln 16611054) each clik anchor has a damaged eyelet with missing silicon.
|