Subject device is a reusable item.However, it cannot be confirmed whether it was reprocessed.Subject device was returned for evaluation.Visual inspection confirmed the failure mode of ¿dimensional¿.The overall length dimension was found to be beyond the specification limit.Further assessment found the end of the device to have damage consistent with an impact with a hard object.An attempt was made to pull the shaft from the handle by hand.However, the shaft was firmly lodged within the handle and could not be removed.Review of the device history records was performed which confirmed no discrepancies.A complaint history review for the catalog number has not identified additional complaints filed for the manufactured lot.Per the evaluation, no root cause could be determined.At this time, no further investigation will be implemented.(b)(4).
|