It was reported that during use of the device for a cardiopulmonary bypass procedure, the accuracy of temperature reading couldn't be trusted.Previous to software upgrade, the arterial temperature on blood parameter monitor (bpm) was approximately one degree higher than the perfusion system oxygenator outlet temperature reading.The customer always rewarmed their patient based upon bpm reading to keep arterial temperature below 37 degrees celsius and have a known element of safety.Now the bpm unit reads approximately one degree lower than the perfusion system temperature reading, meaning we cannot trust which temperature is now accurate.Per the clinical summary on (b)(6) 2015: the australian subsidiary has reported that one of their customers (b)(6) hospital has noticed a number of issues since software version 1.69 has been installed in their units.One observation is the shunt sensor temperature is measuring lower (compared to oxygenator outlet) than in previous software.Previously, the bpm shunt sensor temperature measure was higher than the oxygenator.This issue has been reported by some other centers that have had the 1.69 upgrade.The instructions for use (ifu) is quite clear on the temperature difference between the oxygenator outlet and the shunt sensor, as the shunt is a distance away and the operating room (o/r) is quite cool.In addition, due to small shunt tubing, there is a loss of temperature as the shunt is downstream from the oxygenator.The algorithm is changed for temperature and the accuracy is tighter with the new software.Their observations, regarding temperature, do not indicate a malfunction of the shunt but is a result of improved accuracy for temperature.Manufacturer's global marketing has spoken to the australia subsidiary who will work with the customer to better explain the difference in the how the unit behaves with the new software.Cases have been completed successfully, without delay and without associated blood loss.There has been no harm observed.
|