• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CARDIOPULMONARY BYPASS PUMP SPEED CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CARDIOPULMONARY BYPASS PUMP SPEED CONTROL Back to Search Results
Model Number 60-00-60
Device Problems Device Inoperable (1663); Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/02/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the centrifugal pump touch screen was unresponsive during a procedure.There was no report of patient injury.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the scp centrifugal pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).A sorin group field service representative was dispatched to the facility to investigate.The service representative was unable to reproduce the reported fault, however a data log analysis confirmed the reported issue.Residue was identified near the gasket of the touch screen.However, no fluid ingress was found.The touch screen was cleaned and tested and subsequent testing found the device to be functioning according to specification.Sorin group (b)(4) has not been made aware of any additional issues with the reported unit.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.Evaluated on site by sorin service rep.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CARDIOPULMONARY BYPASS PUMP SPEED CONTROL
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munich D-809 39
GM  D-80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5001247
MDR Text Key24632536
Report Number9611109-2015-00288
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Followup
Report Date 01/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-00-60
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-