Sorin group (b)(4) manufactures the scp centrifugal pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).A sorin group field service representative was dispatched to the facility to investigate.The service representative was unable to reproduce the reported fault, however a data log analysis confirmed the reported issue.Residue was identified near the gasket of the touch screen.However, no fluid ingress was found.The touch screen was cleaned and tested and subsequent testing found the device to be functioning according to specification.Sorin group (b)(4) has not been made aware of any additional issues with the reported unit.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.Evaluated on site by sorin service rep.
|