• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 ROLLER PU,P; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 ROLLER PU,P; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems Device Inoperable (1663); Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group received a report that the pump touchscreen was unresponsive during the procedure.There was no report of patient injury.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 roller pump.The event occurred in (b)(6).(b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative tested the device and was unable to reproduce the reported malfunction.Visual inspection identified build up on the gasket of the touch screen that could be caused by cleaning the screen with wipes.The service representative advised the customer to clean the touch screen with a dry cloth only.As a precaution, the touch screen was replaced.After service, it was confirmed by the perfusionist that the system worked within specifications.The customer kept the old touch screen.No further issues have been reported for this unit.As the issue was not reproduced and the customer kept the replaced touch screen, a root cause could not be determined and corrective actions were not identified.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Device evaluated on site by livanova technician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PU,P
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munich D8093 9
GM  D80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5001368
MDR Text Key23165864
Report Number9611109-2015-00287
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Followup
Report Date 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-