ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST Back to Search Results
Model Number 100071
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 07/16/2015
Event Type  malfunction  
Manufacturer Narrative
It is indicated that the product is not returning for evaluation.Since the products associated with the complaint was not returned, a review of in-house testing data was performed.Retain strip testing results met accuracy and repeatability criteria.The product performed as expected and no product deficiencies were observed.A review of the manufacturing records for the lot did not uncover any non-conformances.The lot meets release specification.Although an improper technique was identified, a root cause could not be determined without additional information.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Event Description
The caller alleged a variance between the inratio inr result in comparison to the lab inr result.The results were as follows: (b)(6) inratio=1.0, poc inr=3.9.The patient self tester's therapeutic range: 2.0-3.0.Multiple finger sticks were performed on the same finger.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INRATIO PT/INR TEST STRIPS
Type of Device
PROTHROMBIN TIME TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer (Section G)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer Contact
ya-ling king
9975 summers ridge rd.
san diego, CA 92121
8588052084
MDR Report Key5001955
MDR Text Key24197476
Report Number2027969-2015-00585
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Reporter Occupation Patient
Report Date 07/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number100071
Device Lot Number368477A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/13/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-