• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: POSEY COMPANY POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

POSEY COMPANY POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS Back to Search Results
Model Number 8070
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Results: evaluation of the returned product found an open zipper slider on the left side window of a patient access area which would not allow proper element closure.Delamination of the mattress envelope on the right side panel was also noted.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not put into use with a patient and should be returned to posey for servicing.If the zipper slider body is bent open (i.E., the mouth is wider than manufacturing specifications) it can prevent the slider from properly engaging the zipper teeth.In areas where the slider engages the teeth, the zipper is secure and cannot be opened.However, if the teeth do not engage it can potentially leave an unsecured area.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.In order for an open slider to potentially contribute to patient egress, the following must occur: the slider does not engage the teeth, the caregiver does not properly check the zipper prior to leaving the patient unattended, and the patient identifies the unsecured area and exits the bed.Following the ifu and standard servicing protocols, the user can identify any open sliders prior to use and return the bed for repair.In this case, there was no impact or consequence to the patient and the canopy was returned for servicing when the zipper issues were identified.Although the cause of the open slider cannot be confirmed, the canopy was approximately 40 months old and there was evidence of improper laundering (material delamination).Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.Note: instructions for use states: never use the posey bed if there is damage to the canopy, access panels or zippers.A failure to heed this warning may result in patient escape or unassisted bed exit, which may lead to serious injury or death from a fall.Always check the canopy and the zippers before leaving the patient alone to help reduce the risk of a fall or unassisted bed exit.Never rip the panels open, as this will damage the zipper slider, preventing the zipper from closing completely.Before leaving the patient alone, apply pressure with your hands along the entire length of the zipper to make sure the panel is completely closed and that there are no gaps or openings along the entire length of each zipper.(b)(4).
 
Event Description
Customer reported damage to the zipper located on the left side panel.No patient incident or injury was reported and the customer did not know when the issue was discovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
POSEY COMPANY
5635 peck road
arcadia CA 91006 0020
Manufacturer (Section G)
POSEY, S. DE R.L. DE C.V.
ave. ferrocarril no. 16901. bo
colonia rio tijuana, 3ra. etap
tijuana 22664
MX   22664
Manufacturer Contact
william hincy
posey company
5635 peck road
arcadia, CA 91006
6264433143
MDR Report Key5005662
MDR Text Key23123342
Report Number2020362-2015-00060
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK103817
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Report Date 08/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8070
Device Catalogue Number8070
Device Lot Number2012040905
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-