Following a pacemaker generator change out, it was noticed by the st.Jude rep that the pacemaker was not pacing.It was determined by x-ray that the ventricular lead was not placed (screwed) in the generator correctly.We had to re-prep the patient, set up for another procedure and reopen the pocket to screw the lead into the generator.This caused prolonged case.After closing the pocket when the pacemaker was being re-interrogated, it was found that the ventricular lead had a very high impedance and further evaluation revealed that the pin connecting the lead to the generator was not making contact.At this point, the pacemaker company representative mentioned that some pacemakers may have been coming from the factory with the "pin" clamping screw not set correctly.After some discussion, the pocket was reopened and that turned out to be the case.The pin was adjusted and re-tightened.Thereafter, excellent measurements were obtained.The generator was replaced in the pocket, irrigated again with several hundred cc's of triple antibiotic solution and closed.Manufacturer response for pacemaker generator, st.Jude 5386 (per site reporter).Rep on site.
|